Agilent Technologies Special Turbo-V 1K-G Manual De Instrucciones página 234

Ocultar thumbs Ver también para Special Turbo-V 1K-G:
Tabla de contenido
17
Technical Information
Window-Meanings
N.
108
109
110
111
120
121
122
123
124
125
126
130
200
201
234/258
Read/
Data
Description
Write
Type
R/W
N
Baud rate
W
L
Pump life/ cycle time/
cycle number reset
R/W
L
Interlock type
(default = 1)
R/W
L
Analog output type:
output voltage signal
proportional to
frequency or power
R/W
N
Rotational frequency
setting (Hz)
R/W
N
Maximum rotational
frequency in Hz (active
only in Stop condition)
R/W
L
Set vent valve on/off
(on = closed)
Reserved to Agilent service
R/W
L
Set the vent valve
operation
R/W
N
Vent valve opening
delay
(expressed in 0.2 sec)
Reserved to Agilent service
R
N
Pump current in mA
dc
R
N
Pump voltage in Vdc
Special Turbo-V 1K-G User Manual / 87-901-006-01 (A)
Admitted Values
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
To reset write '1'
Impulse = 0
Continuous = 1
0 = frequency
1 = power
(default = 0)
500 to 750
(default = 750)
500 to 750
(default = 750)
On = 1
Off = 0
(default = 1)
Automatic = 0 (see note 1.)
On command = 1 (see note 2.)
0 to 65535
(corresponding to 0 to 13107
sec)
Tabla de contenido
loading

Este manual también es adecuado para:

869-8961969-8978m002

Tabla de contenido