SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
NOTE: If the outer ring (sleeve) is in the UP position (GREEN dots not visible),
the connector fillport will not be able to accept the cylinder fillport. Pushing
DOWN momentarily will reset the connector fillport (GREEN dots visible) to
accept the cylinder fillport.
PUSH DOWN
GREEN Dots
6.
Grasp the cylinder/regulator assembly in the area behind the cylinder
gauge (FIGURE 4.4).
7.
Position the cylinder in the compressor cradle (FIGURE 4.4).
8.
Align the cylinder fillport with the connector fillport (FIGURE 4.4).
9.
Pull UP on the outer ring (sleeve) of the connector fillport while
pushing DOWN on the cylinder/regulator assembly to couple the
cylinder fillport into the connector fillport (FIGURE 4.4).
NOTE: The cylinder is properly connected when an audible "click" is heard.
HomeFill
®
Outer Ring (Sleeve)
FIGURE 4.3 Resetting Connector Fillport
BEFORE
COUPLING
CYLINDER,
PUSH DOWN
ON SLEEVE
UNTIL GREEN
DOTS ARE
VISIBLE
16
Connector Fillport
Part No 1167426