SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
SECTION 4—INSPECTION AND
COMPRESSOR INTERFACE
NOTE: Refer to the HomeFill owner's manual for compressor operating
instructions.

Cylinder Prefill Inspection

DANGER
Improper filling or abuse of this cylinder, or failure to
follow these safety instructions may cause serious bodily
injury or death. High pressure oxidizing gas vigorously
accelerates combustion.
DO NOT alter or modify cylinder or related components.
DO NOT use oil or grease. Keep cylinder, cylinder valve,
all equipment and connections free of oil and grease to
avoid violent ignition resulting in serious bodily injury or
death.
WARNING
For medical applications, use only as directed by a licensed
practitioner. Uninterrupted use of high concentrations of oxygen
over a long duration, without monitoring its effect on oxygen
content of arterial blood, may be harmful.
Oxygen cylinders filled by the HomeFill shall be used for personal
use only. "Not to be filled for resale or use by professional users".
This cylinder is to be used with the HomeFill system. This cylinder
contains 93±3% oxygen produced by an oxygen concentration
using the pressure swing adsorption process.
DO NOT handle cylinder or use contents until you are
professionally trained, including emergency procedures.