Abbott FreeGo Manual De Uso página 57

Bomba para alimentación enteral
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

DIAL /
BUTTON
PAUSA
(HOLD)
VOL ADM
(VOL FED)
VOL CERO
(CLEAR VOL)
FUNCTION
Primary purpose is to remind patient that the pump is temporarily
paused. Pump will not automatically restart the feeding after the hold
time has elapsed. To restart the feeding, user must turn dial back to
MARCHA (RUN).
To programme hold time, use Up and Down Arrow Buttons. Keeping an
Arrow Button depressed will make the numerals scroll.
Display will read "MINUTOS DE PAUSA" ("HOLD MINUTES"), along
with the current remaining hold time (from 1 to 90 minutes). Default is
5 minutes, and the pump's hold time resets back to this default value
every time the dial is switched away from PAUSA (HOLD).
Primary purpose is to display the accumulated volume of feeding
solution that has been delivered.
When this number matches the DOSIS (SET DOSE), feeding will
stop and an audio alarm will sound along with a visual display alarm
(reading "FIN DOSIS ADMINISTRADA" ("DOSE DONE")).
Once upper limit (9999 mL) has been reached, the display will show
"- - - -" and is not accompanied by alarm. Therefore, it is critical to
perform VOL CERO (CLEAR VOL) function (below) prior to beginning
feeding.
Display will read "VOL ADMINISTRADO" ("VOLUME FED"), along with
the current accumulated fluid volume that has been delivered (from 1
to 9999 mL).
Primary purpose is to allow the user to clear the accumulated volume of
feeding solution that has been delivered and been stored in the pump's
memory (shown at dial position VOL ADM (VOL FED)).
If lock-out feature is not set, the display will read "BORR EN n S"
("CLEAR IN n SEC"), where n represents the number of seconds
before the accumulated volume delivered is set back to 0 mL. This is
a countdown from 5 seconds, and each second is accompanied by
a beep. When the volume has been cleared, a beep will sound for 2
seconds.
If you do not wish to clear the accumulated volume delivered, turn the
dial away from VOL CERO (CLEAR VOL) before the 5 seconds has
elapsed.
If lock-out feature is set, the display will read "BOMBA BLOQUEADA"
("PUMP LOCKED").
The Up and Down Arrow Buttons are used for increasing or decreasing values,
respectively. They also serve to toggle between different settings (see
44838 Rev. J
below).
Page 55

Hide quick links:

Capítulos

Tabla de contenido
loading

Tabla de contenido