PRIMARY SECONDARY
DISPLAYS
NING
ALIMENTO
(FEED EMPTY)
CONDITION OR
PROBLEM
Air is detected in the
feeding set.
This alarm condition
may be caused by
electromagnetic
interference
(reference the
Guidance and
Manufacturer's
Declaration -
Electromagnetic
immunity table in the
Standards section)
44838 Rev. J
EXPLANATION OR CORRECTIVE
ACTION
If you are just beginning a feeding:
•
Open the door. Gently press the
tubing to ensure it is fully seated in
slots below the cassette. Close the
door firmly.
Or... If there is more food to deliver:
•
Turn dial to PAUSA (HOLD) to stop
the audio alarm
•
Remove feeding set adapter from
your enteral feeding tube
•
Refer to INSTRUCTIONS FOR USE
to prime the air out of the feeding set
•
Securely re-attach the feeding set to
your enteral feeding tube
•
Turn dial to MARCHA (RUN) to begin
feeding
Or... If you are done feeding:
•
Turn dial to OFF/CARGA (OFF/
CHRG). This message is displayed
because pump was programmed to
feed until container is empty (DOSIS
(SET DOSE) at INF DOSIS (infinite
dose)).
•
Move pump away from cellular
phones, amateur radios, AM and
FM radio broadcast, TV broadcast
or other radio frequency (RF)
transmitting devices.
Page 51