Abbott FreeGo Manual De Uso página 49

Bomba para alimentación enteral
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

USING THE NURSE CALL ACCESSORY CABLE (S468):
CAUTION: The pump COMM (communication output) setting must be configured
to NC (for "Nurse Call", Nurse Call Accessory Cable).
NOTE: Pumps with software version V1:01 are not configured for use with the
Nurse Call Accessory Cable (S468). To check the software version of your pump,
from the OFF/CHRG position, turn the pump dial to SET RATE. During the system
Self-Check Procedure the pump display will show the software version. If your
pump has software version V1:01, contact your local Abbott representative to have
the pump upgraded.
To configure your pump for use with the Nurse Call Accessory Cable:
Press the Options Button
until COMM is displayed on the pump display.
Then use the Up and Down Arrow Buttons until NC is displayed.
Pump must be plugged into live AC/Mains power in order for Nurse Call Accessory
Cable to work. Refer to the drawing below for cable and charger connections.
44838 Rev. J
Page 47