This menu contains parameters that determine the operating mode of
the inputs and outputs of the controller.
START MODE: it distinguish three ways to start/stop the pump
according to following description:
-
-
INTERLOCK TYPE: selects the interlock mode on I/O connector:
-
-
ANALOG OUTPUT TYPE: referred to the ANALOG OUTPUT SIGNAL
on I/O connector; it can be related to one of following quantity:
-
-
-
SPEED DISPLAY: the rotational speed can be displayed in
-
-
LCD CONTRAST, increases/decreases the contrast on LCD display.
Turbo-V 551 Rack Controller User Manual / 87-901-031-01
START INTERLOCK, the controller checks for interlock mode;
see interlock type description for further details.
START/STOP LEVEL; if selected, close the START/STOP contact
on I/O connector (1-9 and 2-15) to start the pump and close the
INTERLOCK signal (3-9 and 4-15) to stop the pump.
CONTINOUS: if enabled, the unit wants the contact on I/O
connector to be closed; should be the contact open, error
condition WAITING FOR INTERLOCK is displayed;
IMPULSIVE: if enabled, the unit wants the interlock to be closed
only at start command; should be removed the contact after start
command or during normal operation, no error is displayed.
FREQUENCY, default value
POWER
TEMPERATURE
KRPM
Hz
Technical Information
INPUT/OUTPUT Menu
(A)
17
263/288
loading