SERVICE INFORMATION
The height gage has been properly assembled and calibrated at the factory. There are
no adjustments that can be made without affecting the accuracy of this instrument,
which may incur costly repairs. If the gage has been disassembled, repair charges will
be required. If a malfunction occurs, consult the following checklist to insure that the
problems are not due to improper operation of the gage.
LOSS OF ACCURACY
1.
Is gage base and surface plate clean?
PROBLEM
Does not count
Error Display/Over
Speed Indicator
Low Battery
2.
Is indicator or scriber securely fastened in place?
3.
Is the indicator or scriber clean and free of debris?
4.
Is accuracy being checked against a calibrated standard?
5.
Is the gage being used in an extremely humid environment?
6.
Is the gage subject to sprays of cutting oils, other coolants and liquids?
If the height gage is still inoperative, return it to us for repair.
Please include a letter of explanation and ship to:
Repair Department
The L.S. Starrett Company
165 Crescent Street
Athol, MA 01331-1915
20
PKG08762-UM3754
WHAT TO CHECK
Is the gage in "HOLD" mode - Check by depressing "HOLD"
button and moving the slide.
The word "ERR04" is displayed when the gage slide is moved
at excessive speed. This feature will cancel the display rather
than show an incorrect reading.
The display will show a battery symbol when the batteries are
getting weak. Replace with two new CR2032 batteries.
See "Installing the Batteries" on page 13