WARNING
You must fill the empty space of the newly
inserted head-set after connecting the transfer-
set. Failure to do so results in missed insulin.
If the coupling does not click noticeably and au-
dibly together, use a new head-set and a new
transfer-set.
Do not use any tools to screw the luer-lock con-
nector into place.
Filling volume transfer-set:
12"/ 30 cm: approx. 6 units* approx. 60 μl
24"/ 60 cm: approx. 11 units* approx. 110 μl
31"/ 80 cm: approx. 14 units* approx. 140 μl
43"/110 cm: approx. 18 units* approx. 180 μl
12
Filling volume head-set:
approx. 1.0 units* approx. 10 μl
*U 100 Insulin
WARNING
Change head-set every 2–3 days (max. 72
hours). Otherwise you risk insulin resistance or
inflammation of the infusion site. Never use the
transfer-set for more than 6 days.