Table 1 RS232 commands (continued)
Command Function
RDC
Sets the calibration to the factory default.
CLN
Starts a cleaning at the end of the current cycle.
CTM
Gets the auto calibration settings.
CTM=4,x
Set the auto calibrations to occur every 12 hours. Options: x =
0–23 (24-hour clock)
CTM=5,x
Set the auto calibrations to occur daily. Options: x = 0–23 (24-hour
clock)
CTM=x,y
Set the auto calibrations to occur weekly. Options: x = 6–12 where
Monday = 6 and Sunday = 12, y = 0–23 (24-hour clock)
CLI=x
Gets the auto cleaning setting. Add "=x" to set the auto cleaning
setting. Options: x = 1 (6 hours), 2 (12 hours) or 3 (24 hours)
PMA
Starts priming all the reagents and standards.
PMR
Starts priming all the reagents at the end of the current cycle.
PMS
Starts priming all the standards at the end of the current cycle.
SEQ
Gets "SEQUENCE=12 ....1" which represents the order in which the
sample streams are measured.
SEQ=12
Sets the sample stream sequence. The numbers after "=" represent
the order in which the sample streams are measured (maximum of
8 characters).
RID
Gets the analyzer version (e.g., Alkalinity V1.0 C1.0).
RAR
Gets the latest alarms/warnings (maximum of 10) since the last
RAR command was sent. Enabled only if the SIOx setting is set to
Queried.
RMR
Gets the latest measurement readings for all cycles (grab, normal
and calibration readings) (maximum of 10). Enabled only if the SIOx
setting is set to Queried.
RFL
Sets all the standard, reagent and cleaning solution levels to 100%.
12 English
Table 1 RS232 commands (continued)
Command Function
RG1=xxx
Gets the fluid level of the reagent in percent (e.g., RG1 = Reagent
1). Add "=xxx" to set the fluid level of the reagent in percent.
Options: x = 0–100
SD1=xxx
Gets the fluid level of the standard solution in percent (e.g., SD1 =
Standard 1). Add "=xxx" to set the fluid level of the standard
solution in percent. Options: x = 0–100
WSH=xxx
Gets the fluid level of the cleaning solution in percent. Add "=xxx" to
set the fluid level of the cleaning solution in percent. Options: x =
0–100
ALL
Gets all the analyzer settings.
RIN=xxx
Gets the analyzer name. Add "=xxx" to set the analyzer name
(maximum of 12 characters after the equal sign). All letters will be
converted to upper case.
RCA=x
Sets the output value of the recorder output connection. Options: x
= F, H or Z. F = full-scale output (20 mA), H = half scale output
RCB=x
(12 mA) and Z = zero scale output (4 mA). To stop the test, send
just the first three letters.
RLA=x
Sets the state of the relay. Options: x = 0 or 1 (0 = de-energized,
1 = energized). To stop the test, send just the first three letters.
RLB=x
RVN
Gets the version numbers of the analyzer and the modules. For
example, Hardness V1.0 C1.0, Valve module: VLV0103, Burette
module: PMP0102, ...
TAL
Gets all the data in the alarm/warning logs.
WAL
Erases the data in the alarm/warning logs.