temperature measurements, please contact HoMedics
Consumer Relations at 1-800-466-3342.
TROUBLESHOOTING/ERROR CODES
Condition
Possible Cause
Consistent low
Temperature
temperature
reading is below
readings ("Lo")
measurement range
(below 93.2˚F/
34.0°C).
The probe is not
positioned properly
in ear.
The waterproof probe
is dirty.
Consistent high
Temperature
temperature
reading is above
readings ("Hi")
measurement range
(above 109.4˚F/
43.0°C).
Consistent error
Operating
messages ("Err")
temperature
is not within the
permissible range for
measurement.
Low battery warning
Battery power
is low.
Low battery warning
Battery power is
with "Lo" symbol
too low for the
thermometer to
function.
Corrective Action
Product Design-
Measurement range is 93.2˚F
(34.0°C) to 109.4˚F (43.0°C).
Contact your physician if you have
or suspect any medical problems.
Make sure probe is properly
positioned against the opening of
the ear canal
Clean the probe with a cotton
swab moistened with alcohol.
Product Design-
Measurement range is 93.2˚F
(34°C) to 109.4˚F (43°C). Contact
your physician if you have or
suspect any medical problems.
Product Design-
Operating temperature is
60.8˚F (16°C) to 104˚F (40°C).
Replace battery (see "Inserting
and Replacing Batteries").
Replace battery
immediately (see "Inserting and
Replacing Batteries").
16
PRODUCT SPECIFICATIONS
Model Number
Model Number TE-100
Power Source
DC 3 Volt (lithium battery CR2032)
Battery Life
1,700 Measurements
Sensor
Thermopile
Temperature Indication
In 4 digits, every 0.1˚F or ˚C
+/- 0.4˚F/0.2°C between 96.8˚F -
102.2˚F (36.0°C - 39.0°C)
Accuracy
+/- 0.5˚F/0.3°C less than 96.8˚F (36.0°C)
or greater than 102.2˚F (39.0°C)
Measurement Range
93.2˚F - 109.4˚F (34˚C - 43˚C)
Weight
0.10 lbs (with battery)
60.8˚F - 104˚F (16˚C - 40˚C)
Operating Temperature
Relative humidity up to 95%
Operating Humidity
(non-condensing)
-4˚F - 122˚F (-20˚C - 50˚C)
Storage Temperature
Relative humidity up to 95%
Storage Humidity
(non-condensing)
Ear Thermometer & Battery, Storage Case,
Contents
Instruction Manual
NOTE: These specifications are subject to change without notice. This ear
thermometer meets requirements established in ASTM Standard E1965-98.
17