02.2 Û ULTRASONIC BATH OR AUTOMATED CLEANING AND
THERMODISINFECTION
ULTRASONIC BATH
3,8 bar
10 sec
02 Û ULTRASONIC BATH
Place the insert tip/s in a mesh bottom
1
stainless steel instrument tray;
Place the tray into the ultrasonic cleaner
and ensure that the instruments in the
bath, at 40°C, are submerged by the
enzymatic solution;
Sonicate for at least 10 minutes;
Remove the insert from the ultrasonic tank
2
and rinse with distilled water;
Scrub the inner and outer surfaces of the
3
insert with a clean soft bristled nylon brush
under running tap water;
Rinse with distilled water;
Flush the internal channel of the insert
4
with distilled water injected under pressure
(3,8 bar) for at least 10 seconds so as to
remove any residues;
EN