ACA Current (Clamp-on)
40.00A, 400.0A, 600A
40.00A, 400.0A
40.00A, 400.0A
ACA AutoVATM Threshold: 1A AC (40Hz ~ 500Hz only) nominal
Crest Factor:
< 3 : 1 at full scale & < 6 : 1 at half scale for 40.00A, 400.0A & 600A ranges
1. Induced error from adjacent current-carrying conductor: < 0.06A/A
2. Specified accuracy is from 1% to 100% of range and for measurements
made at the jaw center. When the conductor is not positioned at the jaw
center, position errors introduced are:
• Add 1% to specified accuracy for measurements made WITHIN jaw
marking lines (away from jaw opening)
• Add 4% to specified accuracy for measurements made BEYOND jaw
marking lines (toward jaws opening)
PEAK-rms HOLD (ACA & ACV only)
Response: 65ms to >90%
Frequency
5Hz ~ 500Hz
Sensitivity (Sine RMS)
40A range: > 4A
400A range: > 40A
600A: > 400A
600V range: > 30V
Range
600A
500Hz ~ 3.1kHz
600A
Range
50Hz / 60Hz
45Hz ~500Hz
19
Accuracy 1) 2)
1.0% + 5d
2.0% + 5d
2.5% + 5d
2.5% + 5d
3.0% + 5d
Accuracy
0.5%+4d