Operating Instructions S1-E / S1-A
134°C
65 82 832 D3643.201.01.02.02
7.5
Sterilizing
✔ The instrument is clean and disinfected.
Action steps_instruments
274°F
✔ Instrument maintenance is complete.
✔ The instrument can be sterilized in packaging suitable for
sterilization and storage: paper/plastic composite packaging
according to ISO 11607.
➢ Sterilize the instrument in the steam sterilizer with saturated water
vapor.
Unit pressure in bar and psi ww
Overpressure:
Temperature:
Holding time:
2.04 bar (29.59 psi)
134° C (274° F)
3 min