OPERATION AND USE

Knowing how the plant operates will make the surgery staff
self-sufficient and self-confident.

Dry and semi-dry systems

(draw. 29-30a-31-33a-33b-34a pag. 85-86-87-88-89)
Once the main switch on the feeding line and the control panel one are turned on (I =
on / 0 = off), lifting up a terminal A, draw. 30a, from its seat will start the aspiration: the
microswitch of the tip support, with a 24 V closing signal, controls the opening of the
shutting valve C and starts up the aspiration unit D. Through the operative tip E and the
lifted hose liquids and air, called in by the negative pressure, flow into the manifold B,
where they are submitted to a first filtration. In a dry plant, after passing through the
manifold, liquids get to the Mini-Separator F, where they are drained. In dry systems
only air flows after the Mini-Separator, whilst in semi-dry systems fluids get to the
Separator Tank G draw. 29 (or the centrifuge B, draw. 33a-33b-34a in small semi-dry
systems) in the engine room where they are drained too. In both types of system air
gets to the aspiration units and is exhausted outside, H draw. 29.
If a signal from the Mini-Separator indicates that the maximum level has been reached,
the shutting valve of the involved surgery will close. In semi-dry systems, the Separator-
Tank is fitted with the overflow device that might be activated during washing
operations. The eventual presence of an amalgam separator fitted together with the
Mini-Separator (Hydroseparator) in the dental unit water-unit (dry systems) or installed
in the engine room for semi-dry systems (Hydrocyclone) does not affect or modify the
operation of the plants. The reader can find a more complete and exhaustive description
of the Hydroseparator operation on the specific manual.

MAINTENANCE

This section of the instruction manual, that we would like to
be known also by dental engineers, is particularly meant for
the surgery staff
(draw. 14b-16-17-18-19-29-30a-31 pag. 15-17-85-86)
Before every cleaning or maintenance operation on the aspiration plant, wear
disposable gloves, mask, goggles and overall, carry out a careful sanitizing washing
with Puli-Jet and water, draw. 19, if necessary repeat the operation several times; the
aim is to obtain a good cleaning of the internal components, particularly of the parts
involved in the maintenance operation and to abate as far as possible the bacterial level.
Before starting maintenance operations on moving or under tension parts, cut off
power. To clean and sanitize internally the aspiration plant act as follows, draw. 19:
prepare the Puli-Jet solution following the direction printed on the bottle label; insert the
tip no 17 in the large terminal by means of the adapter no 16; insert the tip no 10 in the
small terminal; dip one tip at a time into the solution; aspirate it without letting any air
enter into the plant; after some seconds lift rapidly the tip upwards and aspirate air only.
Repeat this procedure some times; it will create the turbulence needed to clean all pipes
and components that get in touch with the aspirated liquids. Puli-Jet dissolves mucus
and blood, its special detergent action enchances the antimicrobial factor obtaining a
sanitizing and deodorizing effect. If it is used systematically and regularly, Puli-Jet
assures a good functioning of the plant, prevents foam build-up and formation
16