ELECTRICAL CONNECTIONS
The box receives several cables, i.e. the grey cable which connects the foot-control cable to the box, the black four-wire
cable for serial connection with the dental unit, and the white three-wire cable (brown, blue, and yellow/green) for chair
mains supply.
The following ones are coming from the dental unit:
one three-wire (blue, brown, yellow-green) black cable, one three-wire (blue, brown, yellow-green) white cable, one
more white (two-wire) and one more black one (two-wire), a grey cable for dental unit connection to foot-control
extension, a four-wire black cable for serial connection to the chair, and one more black cable ending by two faston
connections.
The safety connectors (a black and a white one) have ten terminals.
F1
four-wire cable for
serial
connection to chair
four-wire cable for
serial connection to
dental unit
white cable to dental unit
Plug both fast-on wires to the main solenoid valve for air and water (24 V).
Connect the foot-control extension cable (1) to the foot-control cable (2).
Connect the black four-wire cable (3) for chair serial connection to the black four-wire cable (4) from the dental unit.
The wires of the white cable (to switch the suction pump control circuit) must be connected to the first two brown contacts,
whereas the wires of the black cable must be connected to the other two brown contacts (to switch the circuit of nurse
bell or any other service).
3
4
white cable to chair
P.
1
foot-control
extension cable
foot-control cable
to dental unit
2
white wire
black wire
black cable
220 VAC power mains
C
39
ABSOLUTE
Installation instructions