eurodent ABSOLUTE Guia De Instalacion página 35

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

E3
E4
E5
Into the shaped hole on the parallelogram arm of the
chair first insert the suction hose, then the electric cables
wrapped with tape to the Rilsan hoses (green and white),
finally insert the drain hose. Lead all hoses and cables up
to the connecting box.
The accessories boxes contain a foot-control, the suction
tubings with their proper hoses, an assistant syringe (on
request) with a clip, four screws, a grey cord to connect
the foot-control cable to the connecting box, main solenoid
valves for water and air and safety electrical connectors.
The cord has a squared-plug terminal and a circular-
base terminal: push the first terminal mentioned into the
hose which, inside the base footboard, joints the outside
to the connecting box. Then, pull it until the frame of the
circular base adheres to the frame of the hose. Tighten
the four self-tapping screws in the holes of the frame, so
that they penetrate completely the polyurethane and
fasten properly the cord base.
Now, install the instrument table arm of the dental unit
(however first make the guard, cable and hoses run
through the body), and insert it into the proper pin.
Plug the arm cord into X10 (on right top side), on the low-
tension p.c.b.
Then, connect the white hose to the quick fitting, which
provides modules with air in the right valve, and the
green one to the water distribution unit (for tumbler,
assistant syringe and modules), placed on the chair side.
If the Calbenium mixing unit is installed, connect the
green hose to the solenoid valve on the unit.
If the spray feeding unit is installed, connect the green
hose to the unit control switchboard.
35
ABSOLUTE
Installation instructions
Tabla de contenido
loading

Tabla de contenido