► Disinfecting Procedures
1. Put on non-sterile gloves.
2. Take out one disinfecting wipe from the package and squeeze out
any excess liquid in order to prevent damage to the meter.
3. Wipe all meter's exterior surface display and buttons. Hold the
meter with the test strip slot pointing down and wipe the area around
the test slot but be careful not to allow excess liquid to get inside.
Keep meter wet with disinfection solution contained in the wipe for a
minimum of 2 minutes for Micro-Kill+™ wipes. Follow the instructions
on the package label of disinfecting wipe. Use two or more wipes if
necessary.
4. Remove the wipe. Allow the meter surface to dry completely.
5. Discard the used wipes and never reuse them.
6. Remove and discard gloves in appropriate receptacles and wash
hands.
Each cleaning and disinfection cycle includes a pre-cleaning step with
one wipe and a disinfection step with a second wipe.
This device has been validated to withstand 10,950 cleaning and
disinfection cycles using the recommended disinfecting wipe/
towelette.
The tested number of cycles is estimated by 10 cleaning and
36