Step 2
Press Main button to Mark this Test as a Control
Solution Test.
With "QC" displayed, the meter will store your test
result in memory under "QC". If you press the Main
button again, the "QC" will disappear and this test is no
longer a control solution test.
When doing the control solution test, you have to mark it so that the test
result will not mix with the blood glucose stored in the memory.
Step 3
Apply Control Solution.
Shake the control solution vial thoroughly
before use. Squeeze out a drop and wipe
it off, then squeeze another drop and
place it on the tip of the vial cap.
Hold the meter to move the absorbent
hole of test strip to touch the drop. Once
the confirmation window fills completely,
the meter will begin counting down.
To avoid contaminating the control
solution, do not directly apply
control solution onto a strip.
22
Warning