ClEANING, CARE AND ROUTINE mAINTENANCE

Routine maintenance
Servicing of the internal components inside the cabinet of the eclipse must be conducted by a
SeQual trained and qualified service technician as needed.
preventative maintenance (pm) should be completed on a regular basis. Consult your
provider to arrange Pm schedule. Your home care provider or qualified technician will perform
inspections and maintenance of the compressor, alarms, internal 9-volt battery, battery and
other internal parts only as needed.
User Care and Cleaning of the Device
Cannula Replacement: replace your supply tubing and cannula on a regular basis as recom-
mended by your home care provider. Your physician or home care provider will provide you with
cleaning, disinfection and replacement information.
Air Inlet Filter: Ambient air is drawn into the device through the air inlet located on the top
rear of the device. Cleaning the air inlet filter is the most important maintenance activity you will
perform as a user, cleaning the air inlet filter will keep your eclipse performing properly. Check
this air inlet filter daily and clean it at least once a week using the following procedure:
1. remove the air inlet filter;
2. Wash the filter in warm tap water using a
mild soap detergent solution;
3. rinse the filter thoroughly with clear tap
water and squeeze out the excess water;
4. Allow the filter to air dry;
5. reinsert the filter in the cabinet;
Cleaning tip - keep a clean second filter as a replacement for use when one filter is drying.
CAUTION:
THE AIR INLET FILTER SHOULD BE FREE OF LIQUID WATER AND DRY TO THE TOUCH BEFORE
REINSTALLATION. MOISTURE IN THE FILTER MAY IMPAIR PROPER OPERATION OF THE ECLIPSE.
NOTE THAT THE SERVICE PORT IS EXPOSED (WHEN THE AIR INLET FILTER IS REMOVED). DO NOT
ATTACH ANY CABLES TO THIS SERVICE PORT. THE PART IS TO BE USED BY A QUALIFIED SERVICE
TECHNICIAN ONLY.
DO NOT PLACE FOREIGN OBjECTS INTO OR THROUGH THE VENT HOLES OF THE AIR INLET OR
EXHAUST VENT.