RECOmmENDED OpERATING ENvIRONmENTS
For proper use of the device, the following chart provides important information
concerning the recommended operating environments, or operating conditions,
for proper use of the device.
Operating Temperature
Operating humidity
Transport/Storage Temperature
Electrical
Altitude
placement
Environment
Operating Time
NOTE: When moving the Eclipse from an extreme temperature environment, please
allow time for your device to acclimate to the recommended operating temperature
environment.
WARNING:
OPERATING YOUR ECLIPSE OUTSIDE OF THE RECOMMENDED OPERATING ENVIRONMENT MAY
NEGATIVELY IMPACT DEVICE PERFORMANCE, MAY CAUSE DAMAGE TO THE DEVICE AND WILL
VOID THE WARRANTY.
pROpER plACEmENT OF ThE EClIpSE
Select a location for the device that avoids the intake of smoke, fumes and pollutants. Correct
placement of the device should allow intake of air through the air inlet filter at the top, rear of
the cabinet and allow exhaust air to freely leave the exhaust vent at the bottom of the device.
Place the device such that the alarms can be heard.
Position the oxygen supply tubing in such a way that it does not kink or occlude.
DO NOT locate the device near flammable materials or cleaning agents or in the direct path of
any heat source, such as a stove, range, heat register or car heater.
12
Keep the eclipse at least five (5) feet (1.5 m) from heat sources, sparking objects or open flames.
Personal Ambulatory Oxygen System Users Manual
50º F to 104º F (10º C to 40º C)
10% - 95% at an 82.4° F (28° C) dew point
-4º F to 140º F (-20º C to 60º C)
Humidity: up to 95% Non-Condensing
use no extension cords. use no electrical
outlets controlled by a switch.
0 – 13,123 feet (0 - 4,000 meters)
DO NOT block the air inlet or the exhaust vent.
Place device a minimum of 3 inches (7.5 cm)
away from walls, draperies, furniture, etc.
must be smoke, pollutant, and fume free.
24-hours a day when connected to an external
AC or DC power source.