Tabla de contenido
Idiomas disponibles

Idiomas disponibles

TROUBLESHOOTING

If any abnormality arises during use, please check the following points:
SYMPTOMS
Unit does not turn on when
the START/STOP
button is pushed.
EE measurement error
symbol shown on display
or the blood pressure value
is displayed excessively low
(or high).
E1 error symbol shown on
display
E2 error symbol shown on
display
E3 error symbol shown on
display
EP error symbol shown on
display
Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no circumstance
should you disassemble or attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on the warranty page.
POSSIBLE
CAUSES
Batteries have
run down.
Battery polarities have been
positioned incorrectly.
Cuff has been placed incor-
rectly.
Did you talk or
move during
measurement?
Shaking of the wrist with the
cuff on.
Air circuit
abnormality.
Inflation pressure exceeding
300 mmHg.
Error determining measure-
ment data.
System error.
33
CORRECTION
Replace them
with two new "AAA" alkaline
batteries.
Re-insert the
batteries in the
correct positions.
Wrap the cuff properly so that it
is positioned correctly. Measure
again.
Keep wrist steady during
measurement. Measure again.
Measure again.
Switch the unit off, then measure
again.
Measure again.
Measure again.

Capítulos

Tabla de contenido
loading

Tabla de contenido