The tables below provide information on the EMC characteristics (Electromagnetic Compat-
ibility) of this electromedical appliance.
Guide and declaration of the manufacturer – Electromagnetic Emissions
The Corsia aerosol can be used in the following electromagnetic environment. The Cus-
tomer and/or the user of the Corsia aerosol shall make sure that the appliance is used in
such environment.
Emission test
Radiated emissions /
Conductions CISPR11
Radiated emissions /
Conductions CISPR11
Harmonic currents EN
61000-3-2
Voltage fluctuations /
flicker EN 61000-3-3
Guide and declaration of the manufacturer – Electromagnetic Immunity
The Corsia aerosol can be used in the following electromagnetic environment. The Cus-
tomer and/or the user of the Corsia aerosol shall make sure that the appliance is used in
such environment.
Immunity test
Electrostatic
discharge (ESD)
EN 61000-4-2
Fast transient / burst
EN 61000-4-4
Surge EN 61000-4-5 ± 0.5kV e± 1kV
Voltage dips, short
outages and
voltage variations
EN 61000-4-11
Conformity
Group 1
Class [B]
Class [A]
Compliant
Standard of proof
± 8kV contact
± 15kV air
± 2kV power supply
± 1kV signal cables
differential mode
5% U T (>95% dip
in U T ) per 0.5 cycle
40% U T (60% dip
in U T ) per 5 cycles
70% U T (30% dip
in U T ) per 25 cycles
<5% U T (>95% dip
in U T ) per 5 s
ENGLISH
Guide to the electromagnetic
environment
The Corsia aerosol uses RF energy only for
Internal function. Therefore, its RF emissions
are very low and do not cause any interfer-
ence with other nearby electronic applianc-
es.
The Corsia aerosol is designed for use in
any environment, including households and
those directly connected to the public power
distribution grid which supplies power to en-
vironments intended for domestic use.
Level of
Guide to the electromag-
conformity
netic environment
The appliance
Floors should be in wood,
does not alter
cement or ceramics. If floors
its status
are covered by synthetic ma-
terial, the relative humidity
should be at least 30%.
The appliance
The power supply should be
does not alter
that of a typical commercial
its status
premise or hospital.
The appliance
The power supply should be
does not alter
that of a typical commercial
its status
premise or hospital.
--
The power supply should be
that of a typical commercial
premise or hospital.
If the user of the Corsia aer-
osol requires the appliance to
work continuously, it is rec-
ommended to use it with an
uninterruptible power supply.