Metrolab PT2026 Manual De Instalación Y Seguridad página 11

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

PT2026 Installation and Safety Manual
8
USB
9
Auxiliary
 To avoid potentially dangerous equipment failures, only Metrolab
peripherals, designed for the purpose, should be connected to the Probe
connector.
 To avoid electric shock, only limited-energy circuits (in the sense of EN
61010-1) that conform to the Safety Extra Low Voltage standard (IEC
60364-4-41) should be connected to the Trigger, Clock and Auxiliary
connectors.
 Treat the probe connectors with care. They have very tight tolerances, and
crushing them, or even dropping them on the floor, may deform the shell,
making it difficult or impossible to plug in the probe.
 When plugging in the probe, make sure the connector is perfectly aligned
with the socket; otherwise it will jam.
 To unplug the probe, you must grasp the sleeve and pull straight back to
release the locking mechanism. Never pull on the cable.
 To unplug the female plug of the MUX cable from a MUX6026 multiplexer,
press the ring at the base of the socket to release the locking mechanism.
v 1.0 r 4.0 – 02/2017
USB 2.0, Standard B connector
Mini DIN 5:
1. On/Off (+5 V): short to ground is like pressing
power switch;
2. NMR signal output (±5 V);
3. Modulation output (±9 V, continuous-wave
only);
4. Sync output (0–5 V, pulsed-wave only);
5. Ground.
CAUTION
!
7
Copyright © 2017 Metrolab Technology SA
www.metrolab.com
Tabla de contenido
loading

Productos relacionados para Metrolab PT2026

Tabla de contenido