EXFO FTB-500 Guía Del Usario página 438

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

SCPI-Based Errors
Error
Description
Number
–363
"Input buffer overrun"
–365
"Time out error"
–400
"Query error"
–410
"Query INTERRUPTED" [Indicates that a condition causing an
–420
"Query
UNTERMINATED"
–430
"Query DEADLOCKED" [Indicates that a condition causing an
–440
"Query
UNTERMINATED after
indefinite response"
–500
"Power on"
–600
"User request"
430
Probable Cause
[Software or hardware input buffer on serial port
overflows with data caused by improper or
nonexistent pacing.]
[This is a generic device-dependent error.]
[This is the generic query error for devices that
cannot detect more specific errors. This code
indicates only that a Query Error as defined in IEEE
488.2, 11.5.1.1.7 and 6.3 has occurred.]
INTERRUPTED Query error occurred (see IEEE
488.2, 6.3.2.3); for example, a query followed by
DAB or GET before a response was completely
sent.]
[Indicates that a condition causing an
UNTERMINATED Query error occurred (see IEEE
488.2, 6.3.2.2); for example, the device was
addressed to talk and an incomplete program
message was received.]
DEADLOCKED Query error occurred (see IEEE
488.2, 6.3.1.7); for example, both input buffer and
output buffer are full and the device cannot
continue.]
[Indicates that a query was received in the same
program message after an query requesting an
indefinite response was executed (see IEEE 488.2,
6.5.7.5).]
[The instrument has detected an off to on
transition in its power supply.]
[The instrument has detected the activation of a
user request local control.]
FTB-500
Tabla de contenido
loading