DISASSEMBLY OF PARTS FOR CLEANING
AND STERILIZATION
Before moving on to the cleaning procedures described in the following chapters, disconnect
all the accessories and components of the PIEZOSURGERY
off
03
.
®
1
WARNING: Switch the device
off. Always turn the device off via
the switch, and disconnect the
power supply cable from the power
outlet and from the device console,
before performing the cleaning and
sterilization interventions.
WARNING: Supplementary
grounding cable. If present, disconnect
the supplementary grounding cable
before you perform the cleaning and
sterilization interventions.
2
the irrigation bag support rod from the
device console;
Disconnect the foot pedal from the device:
3
Grab the connector of the foot pedal, press
back;
CAUTION: Do not attempt to
unscrew or turn the connector during the
disconnection: the connector could get
damaged.
CAUTION: When disconnecting the
foot pedal, always and only hold the
connector of the cord. Never actually
pull on the cord itself.
DISASSEMBLY
EN
7