01
INTRODUCTION
Carefully read this manual before proceeding
with the cleaning and sterilization operations,
and always keep it within reach.
IMPORTANT: to prevent harm to persons
or damages to objects, read all the "Safety
precautions"present in the manual with
special care. Depending on their degree
of seriousness, the safety precautions are
WARNING
(always refers to personal injury)
CAUTION
(refers to possible damage to
property)
The purpose of this manual is to make
the operator knowledgeable of the safety
precautions, the installation procedures,
and the instructions for a correct use and
maintenance of the device and its accessories.
Use of this manual for purposes other than
those strictly tied to cleaning and sterilization
of the device is forbidden.
The information and illustrations in this
manual are updated as of the date of issue
reported on the last page.
Mectron is committed to continuously
update its products with possible
In case you uncover discrepancies between
what described in this manual and the device
in your possession, contact your Retailer
or the After-Sales Service of Mectron for
Limitations on the number of cleaning and
sterilization cycles: Cleaning and sterilization
cycles repeated over time have minimum
effects on the instruments examined in this
Manual. The life expectancy of the products is
usually determined by their consumption and
by damages caused from their use.
WARNING: The operators who
perform the cleaning and sterilization
operations must be adequately
protected and trained
WARNING: Infections control.
First use: All new and repaired accessories
are supplied in NON STERILE conditions.
Before use, and after each treatment, they
must be cleaned and sterilised in strict
compliance with the instructions given in
the Cleaning and Sterilization Manual.
01
INTRODUCTION
Subsequent uses: After every treatment,
clean and sterilize all the reusable parts
and accessories, following the instructions
provided in the Cleaning and Sterilization
Manual.
WARNING: The cleaning processes
must be commenced immediately after
each use. Do not allow that contaminated
instruments dry up before starting the
cleaning and sterilization process. To
eliminate organic remains such as blood,
bone, and other, use an enzymatic
detergent with neutral pH (pH7),
immediately after use.
CAUTION: Do not use metallic
brushes or abrasive sponges during the
cleaning processes, because they damage
the surfaces of the treated parts and the
with soft nylon bristles.
CAUTION: The detergent solution
must be completely removed from the
treated parts to prevent accumulation of
chemical residues.
CAUTION: During cleaning procedures
only use detergents with pH 6-9.
CAUTION: If you intend to disinfect,
we recommend that you use water-based
disinfectant solutions with a neutral
pH (pH 7). Alcohol-based disinfectant
solutions and hydrogen peroxide are
contra-indicated, because they can fade
the color and/or damage the plastic
materials. This also holds true for chemical
products such as acetone and alcohol.
Always rinse with sterile water to preserve
the disinfection.
WARNING: Do not use tap water,
unless where explicitly indicated.
WARNING: Reusable accessories
that must be forwarded to an
Authorized Mectron Service Center.
All the parts must be cleaned and
sterilized in accordance to the procedures
described in this Manual, before they are
sent to an Authorized Mectron Service
Center. Incorrectly treated and bio-
contaminated parts will not be accepted
by the Authorized Mectron Service
Centers.