RPB GX4 08-400 Manual Del Usuario página 20

Ocultar thumbs Ver también para GX4 08-400:
Idiomas disponibles
  • ES

Idiomas disponibles

GX4 GAS MONITOR
MAINTENANCE
FIGURE 10.5
After removing the manifold, inspect it from
all sides to make sure all of the holes are
clear and that there is no oil or debris inside
of the tube. DO NOT DISASSEMBLE THE
MANIFOLD. If there is oil or debris inside of
the manifold a replacement manifold (P/N:
08-422) must be used.
FIGURE 10.7
Rotate the manifold into the back of the
monitor so the clip goes into the notches
on the back of the monitor making sure it is
securely clipped in.
20
CONTINUED
FIGURE 10.6
To replace the manifold, insert the tapered
end into the green ring of the push-lock fitting
inside the back of the monitor. Orient the
holes in the manifold towards the sensors.
REPLACE MOUNTING BRACKETS
If the Mounting Brackets (P/N: 08-425)
break, you can replace them by removing the
two screws securing each bracket. Remove
the broken bracket and replace with a new
bracket. Secure the new bracket with the
provided (UNC 10-24 x ¾") screws.
FIGURE 11.1
loading