4.1.2 Setting Up
Risk of instrument damage due to inappropriate operating conditions
An unsuitable operating site or test setup can damage the instrument and connected
devices. Before switching on the instrument, observe the information on appropriate
operating conditions provided in the data sheet. In particular, ensure the following:
●
All fan openings are unobstructed and the airflow perforations are unimpeded. The
minimum distance from the wall is 10 cm.
●
The instrument is dry and shows no sign of condensation.
●
The instrument is positioned as described in the following sections.
●
The ambient temperature does not exceed the range specified in the data sheet.
●
Signal levels at the input connectors are all within the specified ranges.
●
Signal outputs are connected correctly and are not overloaded.
The R&S NRT2 is designed for use under laboratory conditions.
Bench Top Operation
If the R&S NRT2 is operated on a bench top, the surface should be flat. The instrument
can be used in horizontal position. With the two tiltable feet, the angle of the instrument
can be adjusted.
Risk of injuries
If the R&S NRT2 is not set up securely, you or others can be injured.
Place the R&S NRT2 on a stable and level surface. Do not place anything on top of the
R&S NRT2, if the R&S NRT2 is not in a level position.
Getting Started 1430.0573.02 ─ 02
Preparing for Use
Putting into Operation
10