Resetting to default settings

You can reset the measurement parameters and the configuration to the
delivery status separately from one another (initialization).
The following measurement parameters can be reset to the delivery status:
Measuring mode
Slope
Asymmetry
Manual temperature input
The following configuration parameters (InI) can be reset to the delivery
status:
Baud rate
Display during pH calibration
Interval 1
(automatically saved)
Interval 2
(for data transmission)

Reset procedure:

Press <CAL> while pressing the <RUN/ENTER> key.
pH
To reset the measurement parameters, use < > < > to select YES and
cm
confirm with <RUN/ENTER>, or
Sal
Continue to the configuration parameters without resetting (InI) with the
/
K
TP
<RUN/ENTER> key.
After the InI configuration parameters, the instrument switches to the last
active measurement mode.
Resetting to default settings
pH
-59.16 mV/pH
0 mV
25 °C
4800
Buffer nominal value
OFF
OFF
27