Oxygen calibration procedure:
l Keep a suitable air calibration vessel ready.
l Connect the D. O. probe to the measuring instrument.
l Put the D. O. probe into the air calibration vessel.
l Press the <CAL> key. The oxygen calibration mode appears (display
shows O
CAL).
2
l Press <RUN/ENTER>. The AutoRead measurement begins.
l If the measured value is stable, the instrument displays the value of the
relative slope and the sensor evaluation.
l Switch to the measuring mode with <M>.
Resetting to default settings
The following settings are reset to the status on delivery when a reset is per-
formed (initialized):
Note
During a reset, the relative slope for the sensor types CellOx 325 and
DurOx 325 are reset together. They cannot be reset separately.
Reset procedure:
l Press <CAL> while pressing the <RUN/ENTER> key.
l Press <RUN/ENTER> to reset the measurement parameters. The instru-
ment changes to the measuring mode.
Resetting to default settings
7