3. Acquiring PANO Images
3.2.1
PANO Examination Mode (Standard / Right / Left / Front /
Orthogonal)
Normal Patient Positioning
1.
Insert the Normal Bite into the Normal Chinrest and cover it with a Sanitary Vinyl Cover.
Use the Temple Supports Open/Close button on the control panel to widen the
2.
Temple Supports.
3.
Guide the patient to the inside of the equipment.
4.
Use the Column UP/DOWN button or switch option to adjust the height of the
equipment so that the patient's chin reaches the Chinrest.
5.
Guide the patient to stand in the center of the equipment and direct them to remain in
the position outlined below.
Hold the handles tightly.
Press the chest against the equipment.
Keep both feet close to the inside of the base.
Keep both shoulders parallel.
Straighten the Cervical Spine and stand still.
32
Sanitary Vinyl Cover
Normal Bite
 The Sanitary Vinyl Cover is for single use only. It should be replaced
for each patient. Be sure to use the approved vinyl cover.
 Clean the Chinrest and the Bite with ethanol and wipe with a dry
towel before the next patient.
Green 16 User Manual