Handicare SIMPLICITY Manual De Instalación página 31

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 154
Code:
Meaning:
o
Default Eeprom.
Description:
The Eeprom has been reset to its default value
(Flash Failure).
Telephone fault finding action:
Send engineer.
On site fault finding action:
Replace main board.
Code:
Meaning:
r
No charge current.
Description:
Activated.
Telephone fault finding action:
Check that power supply is switched on from mains
supply. If yes send engineer.
On site fault finding action:
Check power supply.
Code:
Meaning:
y
Software error.
Description:
The main control board has a software fault.
Telephone fault finding action:
Send engineer.
On site fault finding action:
Replace main board.
Code:
Meaning:
.8.
N/A
Lift is in "sleep mode" to
conserve battery power.
Description:
Telephone fault finding action:
None.
On site fault finding action:
Press toggle or handset button to wake lift.
Code:
P
P
Description:
Powered swivel did not respond.
Telephone fault finding action:
NOT YET ACTIVE.
On site fault finding action:
NOT YET ACTIVE.
Code:
U
U
Description:
Faulty power supply.
Telephone fault finding action:
Send engineer.
On site fault finding action:
Replace power supply.
Code:
N/A
Description:
N/A.
Telephone fault finding action:
Send engineer with a service exchange power pack.
On site fault finding action:
Replace power pack.
Note: If the display is showing
anything but the diagnostic codes
listed it requires resetting.
Drive the lift off the charge contacts.
Turn the main switch on the unit off
for 10 seconds and then back on.
HANDICARE SIMPLICITY INSTALLATION MANUAL 31
Appendix 3
Meaning:
PS no reply.
Meaning:
No float indication V1.19
software and later only.
Meaning:
Over current.
Tabla de contenido
loading

Tabla de contenido