119
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Table 1: Guidance and manufacturer's declaration - electromagnet-
ic emissions
The device is intended for use in the electromagnetic environment spec-
ified below. The user of this device should assure that it is used in such
an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Table 2: Guidance and manufacturer's declaration - electromagnet-
ic immunity
The device is intended for use in the electromagnetic environment spec-
ified below. The user of this device should assure that it is used in such
an environment.
Immunity test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC61000-4-4
Compliance Electromagnetic environment
Group 1
Class B
N/A
N/A
IEC 60601
test level1
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for
input/output lines
This device uses RF energy only for its
internal function. Therefore, its RF emis-
sions are very low and are not likely to
cause any interference in nearby electron-
ic equipment.
This device is suitable for use in all estab-
lishments. It uses internal power supply
and has no connection with power supply
network.
Compliance
Electromagnetic
level
environment
±6 kV contact
Floors should be wood,
±8 kV air
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
N/A
N/A