Electromagnetic Compatibility (EMC) to EN60601-
1:2007
WARNING: use of portable phones or other radio frequency (RF)
emitting equipment near the system may cause unexpected
or
adverse operation
The MicroLoop has been tested to EN60601-1-2:2007, regarding it's ability
to operate in an environment containing other electrical/electronic equipment
(including other medical devices).
The purpose of this testing is to ensure that the MicroLoop is not likely to
adversely affect the normal operation of other such equipment and that
other such equipment is not likely to adversely affect the normal operation
of the MicroLoop.
Despite the testing of the MicroLoop that has been undertaken, normal
operation of the MicroLoop can be affected by other electrical/electronic
equipment and portable and mobile RF communications equipment.
As the MicroLoop is medical equipment, special precautions are needed
regarding EMC (electromagnetic compatibility).
It is important that the MicroLoop is configured and installed/put into
service, in accordance with the instructions/guidance provided herein and is
used only in the configuration as supplied.
Changes or modifications to the MicroLoop may result in increased emissions
or decreased immunity of the MicroLoop in relation to EMC performance.
The MicroLoop should be used only with the accessories (USB cables, mains
adapter, cradle and turbine transducer) supplied (which are referenced in
the accessories section of this manual). None of the MicroLoop cables should
be extended in length by the user.
If any cables are extended by the user or non approved accessories are
used, this may result in an increased level of emissions or decreased level of
immunity, in relation to the MicroLoop's EMC. None of the MicroLoops
accessories should be used with other devices, as this may result in an
increased level of emissions or decreased level of immunity, in relation to
the other device's EMC.
The MicroLoop has an essential performance – the product should continue
to operate correctly. In the unlikely event of a Fast Transient / ESD event
occurring, the device should be reset and located away from the source of
interference.
WARNING:
The MicroLoop should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use with other equipment is necessary,
the MicroLoop and the other equipment should be observed/monitored, to verify
normal operation in the configuration in which it will be used.