Technical data

Measurement principle: oscillometric
Measurement range:
40 to 260 mmHg
40 to 200 pulses/min
Display: LCD
Operating conditions:
ambient temperature: 10 °C – 40 °C
relative humidity: 10 – 85 %
Transport and storage conditions:
Temperature -5 °C – 50 °C
relative humidity: max. 85 %
Power supply:
DC 3V (2 x 1.5 V microcells alkali manganese,
IEC LR 03)
Battery check: indicator symbol in display
Weight: 93 g
Dimensions (W x D x H): 75 mm x 65 mm x 60 mm
Classification: Type BF
Clinical trials:
DIN 58130 testing procedures N6
„sequentielle Vergleichsmessung am selben Arm"
Results:
systematic measurement error:
systolic blood pressure: -1.37 mmHg,
diastolic blood pressure: -1.68 mmHg.
Empirical standard devation:
systolic blood pressure: ± 7.52 mmHg,
diastolic blood pressure: ± 5.53 mmHg.
Accuracy:
maximum measuring deviation of the sleeve
pressure: ± 3 mmHg
Error margin of the pulse frequency measuring
device is ± 5 %
Technical specifications subject to change without
prior notification.