Safety
Basic Construction:
Mode of Operation:
Protection Class:
Output Type:
Dimensions and Weight
Length x Width x Height = 9" x 4.5" x 6.3"
Weight: 5 lbs.
WARRANTY AND REPAIR
The Aaron High Frequency Desiccator is covered under warranty for a period of two years. The handpiece is covered
under warranty for a period of one year or 25 steam autoclave cycles, whichever comes first. The warranty becomes
null and void if damage occurs from incorrect handling or misuse of the product.
For warranty and repair work, please contact Aaron and obtain a Return Goods Authorization number (RGA). Place
the number so that it can be seen on the exterior of the package and ship directly to Aaron. A return without an
RGA may not be accepted.
TROUBLESHOOTING
The Aaron High Frequcncy Desiccator has been designed and manufactured with the utmost safety in mind. The
unit is equipped to automatically detect a malfunction. The following table list error codes, their meaning and
recommended actions to be taken to resolve the error.
Error Code
Description of Error
E1
Activation upon power up
E2
DC Supply Over Voltage
Detection
E3
Pulse Width
E4
Delta Error
E5
Multiple Errors
E6
Temperature Error
If problems persist, the unit should be taken out of service and the manufacturer should be notified. For technical
support or return authorization phone +1-800-537-2790.
TRANSPORT AND STORAGE
Ambient temperature range: -40˚C to 70˚C
Relative humidity: 10% to 100%, non-condensing
Atmospheric pressure: 500hPa to 1060hPa.
8
In accordance with EN 60601-1
Intermittent operation
CLASS I EQUIPMENT
TYPE BF
Recommended Action
• Check handpiece for activation.
• Check footswitch for activation; once the activation is
halted the unit will resolve the error. If the error
persists the handpiece could be malfunctioning and
may need to be replaced.
• Switch unit off and on again.
• Make sure unit is connected to correct power source
for the unit.
• Switch unit off and on again.
• Switch unit off and on again.
• Switch unit off and on again.
• Switch unit off. Allow unit to cool. Switch unit on.
Bovie Medical