5.4 Aligning the scale

6. WEIGHING

6.1 Adjusting the display

1. Position the transport lock as shown in the adjacent
drawing.
2. Attach the transport lock to the lower section of the
housing.
ATTENTION!
Incorrect measurement due to force shunts
Soft floor coverings such as carpets may cause
weights to be measured incorrectly.
– Set the scale up so that only its feet are in
contact with the floor.
Place the scale on a firm, flat surface.
CAUTION!
Bodily injury
Perform a function check before every use as
described in the section "Function check" on
page 38.
To obtain meaningful measured values you must check
before every use that the pointer of the scale reads "0".
1. Ensure that the scale has no load.
2. Check the current position of the pointer.
3. Move the adjusting wheel so that the pointer reads
"0".
– Turn clockwise: Pointer moves counter-
clockwise
– Turn counter-clockwise: Pointer moves
clockwise