ELECTROMAGNETIC COMPATIBILITY (EMC)
• Guidance and manufacturer's declaration—electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and should only be used
in such environments:
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions IEC 61000-3-3
• Guidance and manufacturer's declaration—electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only be used
in such environments:
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short interrup-
tions and volt-
age variations
on power supply
input lines
IEC 61000-4-11
Compliance
Group 1
Class B
Class A
Complies
IEC 60601 test level
± 8 kV contact
± 15 kV air
power supply lines: ± 2 kV
input/output lines: ± 1 kV
line(s) to line(s): ± 1 kV
line(s) to earth: ± 2 kV
100 kHz repetition
frequency
0% UT; 0.5 cycle
At 0°, 45°, 90°, 135° , 180°,
225°, 270° and 315°.
0% UT; 1 cycle and 70%
UT; 25/30 cycles
Single phase: at 0°
0% UT; 300 cycles
Electromagnetic environment – guidance
RF energy is used only to maintain device's operation.
Therefore, its RF emissions are so low that it's not likely to cause
any interference in nearby electronic equipment.
The device is suitable for use in all establishments, including
domestic establishments, and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes.
Compliance level
± 8 kV contact
± 15 kV air
power supply lines: ± 2 kV
line(s) to line(s): ± 1 kV
100 kHz repetition
frequency
0% UT; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°.
0% UT; 1 cycle and 70%
UT; 25/30 cycles
Single phase: at 0°
0% UT; 300 cycles
25
Electromagnetic
environment – guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power qual-
ity should be that of a
typical commercial or
hospital environment.