HoMedics BPW-201 Manual De Usario página 33

Monitor automático de presión sanguínea para la muñeca
Ocultar thumbs Ver también para BPW-201:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

• Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only
be used in such environments:
Immunity
IEC 60601
test
test level
Conducted
3 Vrms
RF
150 kHz to
IEC 61000-
80 MHz
4-6
3 V/m
Radiated
80 MHz to
RF
2.5 GHz
IEC 61000-
4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be pre-
dicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmit-
ters, an electromagnetic site survey should be considered. If the measured field strength in the location
in which the device is used exceeds the applicable RF compliance level above, the device should be
observed to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the device.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Compliance
Electromagnetic environment – guidance
level
Not Ap-
Portable and mobile RF communications equipment
plicable
should be used no closer to any part of the device,
including cables, than the recommended separation
3 V/m
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d = 1.2
d = 2 .3
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey, a should be less than
the compliance level in each frequency range.b
Interference may occur in the vicinity of equipment
marked with the following symbol:
32
80 MHz to 800 MHz
800 MHz to 2.5 GHz

Capítulos

Tabla de contenido
loading

Tabla de contenido