Specifications
Specifications are subject to change without notice.
Specification
Details
Dimensions
18.6 x 7.3 x 3.8 cm (7.32 x 2.87 x 1.5 in.)
Weight
300 g (0.66 lb)
Meter enclosure
IP67
Power requirements
AA Alkaline or rechargeable Nickel Metal Hydride
(internal)
(NiMH) batteries (3); battery life: over 500 hours
Input connectors
MP-5 or MP-8 connector for sensION+ probes
Meter protection class
Class III
Storage temperature
–15 to +65 °C (5 to +149 °F)
Operating temperature
0 to 50 °C (32 to 122 °F)
Operating humidity
80% (non-condensing)
Instrument range
pH: –2.00 to 19.99; ORP: ±1999 mV
Conductivity: 0.01 to 500 mS/cm; TDS: 0 to 500 g/
L; Salinity: 0.0 to 1999 mg/L, 2.0 to 50.0 g/L
DO: 0.00 to 19.99 mg/L and 20.0 to 22.0 mg/L
(25°C), 0.0 to 199.9% and 200 to 250% (25°C)
Resolution
pH: 0.01 pH; ORP: 1 mV (0.1 mV from –199.9 to
199.9 mV)
DO: 0.1% (1% if ≥ 200%); 0.01 mg/L (0.1 mg/L if ≥
20 mg/L)
Conductivity: changes with range (auto-ranging)
Temperature: 0.1 °C (0.1°F)
Specification
Details
Measuring error (± 1 digit)
pH: ≤ 0.01 pH; ORP: ≤ 1 mV
DO: ≤ 0.5 of the measured value
Conductivity/Salinity/TDS: ≤ 0.5 %
Temperature: ≤ 0.2 °C (≤ 0.4 °F)
Reproducibility (± 1 digit)
pH: ± 0.01 pH; ORP ± 1 mV
DO: ≤ 0.2 of the measured value
Conductivity/Salinity/TDS: ± 0.1 %
Temperature ± 0.1 °C (± 0.1 °F)
Reference temperature
Conductivity: 20 or 25 °C (68 or 77 °F) (factory
(RT)
settings: 25 °C (77 °F))
Temperature coefficient
Conductivity: 0.00 to 5.00%/°C (factory settings:
(TC)
2%/ °C)
TDS conversion factor
Conductivity: 0.00 to 4.44 (factory settings: 0.64)
Energy management
Automatic power off after 5 minutes of inactivity
Certification
CE

General information

Revised editions are found on the manufacturer's website.

Safety information

Please read this entire manual before unpacking, setting up or operating
this equipment. Pay attention to all danger and caution statements. Failure
to do so could result in serious injury to the operator or damage to the
equipment.
Make sure that the protection provided by this equipment is not impaired,
do not use or install this equipment in any manner other than that specified
in this manual.

Use of hazard information