doesn't share the same the file list with other functions. Select the location of a file first, then perform ope-
rations (save, read, rename or delete).
The introduction of function keys on File List page:
[SAVE] First, select the location where the file is to be saved, press [SAVE], and a numeric keypad will
pop up. Key in a name for the file, press [ENTER] to confirm, and the current curve information will be
saved. The file name must be less than 8 characters. If a selected serial number already has a file, the new
saves file will replace the old one.
[LOAD] First, select the location where the file is
saved, press [LOAD], the page will jump to Curve
Display, and the curve information which has been read
will show on the screen.
[Rename] First, select the location where the file is
saved, press [Rename], and a numeric keypad will pop
up. Key in a new file name, press [ENTER] to confirm,
and the file will have a new name.
[Delete] First, select the location where the file is
saved, press [Delete], and the selected file will be dele-
ted.
5. Multi-wavelength Measurement
Parameter Setting
Press [Multi WL] from the main function menu on the right of the screen to get into the function set-
tings. The first screen is Parameter Setting page. All function tabs from Page 1 to 4 can be found on the
upper left place of the screen, and they can switch back and force.
Introductions of other function keys on the Parameter Setting page:
[ ] and [ ] Cell holder selection key. Press any arrow,
and the position move one. The initial position is R, and
the rest in sequence are: S1, S2, S3, S4, S5, S6, and S7
[R Position] Press this key, the cell holder returns to
the default position R.
[T% / Abs] Two measurement modes can be switched:
transmittance (T %), and Absorbance (Abs).
[AUTOZERO] Adjust zero baseline.
[TEST] On the completion of setting parameters, starts
multi-wavelength testing.
[PRINT] Print the testing results.
Example
Press the circled areas on the image to set wavelength
and the number of sample cuvettes.
Enter wavelength values from small to large.
Next, get reference (blank) sample (which must be put
onto the reference cell holder R) and testing sample in
the matched quartz cuvettes, and place them separately
on to the cell holder R. Press [AUTO ZERO] for the cell
holder R.
Then press [TEST/STOP] key to start testing.
When the testing is finished, press [PRINT] key to
print out the testing data.