6. Care and Maintenance
6.4 Cleaning the Hydrim M2
To clean outer surfaces and the decal covering the touchscreen, use a soft cloth
moistened with soap and water or a surface disinfectant. Do not use harsh cleaning
chemicals.
If preparing the unit for transport or storage, waste water should be removed
from the sump.
Waste water in the unit may contain biological contaminants. Use a mechanical means
to siphon the contents. Wear disposable rubber gloves. Dispose of absorbent material
according to biological waste disposal regulations.
6.5 Filter and Wash Arm Maintenance
Filter
Filter Maintenance:
Inspect the filter in the bottom of the
chamber daily for debris and clean if
necessary. To clean, remove the filter
rinse under a tap and replace. Ensure
that the filter is firmly locked into position
when replaced.
Wash arms
Wash Arm Maintenance:
If the wash arms are not turning easily,
remove them, rinse under a tap and
reassemble. To remove the upper or
middle arm, unscrew the collar
counterclockwise. To remove the lower
arm unscrew the collar clockwise.
54