Basics
4.6

Configuring the test gas inlet

WARNING
!
The gas concentrations entered must be identical to
the specifications on the gas cylinder used. Incorrect
details will result in faulty measurement results.
NOTICE
i
i
When the test gas concentration is changed, the
corresponding test gas inlet must be reconfigured.
To configure a gas inlet:
1. Select
> Station gas configuration.
An overview of the test gas connections is displayed.
Select the test gas inlet
2. Select desired gas inlet.
The configuration menu will appear.
Select the test gas
inlet
Enter part no.
Enter use-by date
Select test gas
Enter concentration
Further options
Default setting for the bump test tolerance: 20 % (5 % with O
24
◄ Enter batch no.
◄ Create Test gas
component
◄ Delete configuration
◄ Select unit
◄ Set bump test
tolerance
)
2
When using a Dräger gas cylinder:
NOTICE
i
i
Upon entering the part no. of a Dräger gas cylinder, a
gas cylinder level indicator is displayed automatically,
unless this function was deactivated before (see
chapter 4.6.1 on page 25).
1. Enter the part no. of the Dräger gas cylinder.
All the necessary details for the configuration will be filled
in automatically. The batch number and the expiry date can
be entered manually in addition.
NOTICE
i
i
The values entered automatically must be matched
with those indicated on the test gas cylinder. If these
values are not identical, the value indicated on the gas
cylinder is to be considered and a manual correction of
the values must be made in the station.
2. If necessary select Further options and
gas cylinder level indicator.
3. If required, configure other gas inlets in the same way.
When using a gas cylinder from another manufacturer:
1. Create or delete test gas component.
Create a new test gas component with
Delete the current test gas component with
NOTICE
i
i
Deleting all test gas components deletes all
information of the test gas inlet.
2. Select gas.
3. Enter the gas concentration.
4. Select gas units.
5. If required, create additional test gas components.
6. The following information can optionally be given:
Part no. of the gas cylinder
Batch number of test gas cylinder
Expiry date of the gas cylinder
7. If required, enter Further options.
8. For more information see X-dock 5300/6300/6600
Technical Manual.