INSTALLATION

Upon receiving the Ohaus Extreme Shaker, check to ensure that no damage has
occurred during shipment. It is important that any damage that occurred in transport
is detected at the time of unpacking. If you do find such damage the carrier must be
notified immediately.
DO NOT lift the unit by the controller or power cord. After unpacking, place the shaker
on a level bench, table, or surface away from explosive vapors. Ensure that the sur-
face on which the unit is placed will withstand typical heat produced by the unit and
place the unit a minimum of six (6) inches from vertical surfaces. Always place the unit
on a sturdy work surface.
The Extreme Shaker is supplied with a power cord that is inserted into the IEC con-
nector on the back of the unit first, then it can be plugged into a properly grounded
outlet. The 120V unit plugs into a 120 volt, 50/60 Hz source. The 230V unit plugs into
a 230 volt, 50/60 Hz source.
MAINTENANCE & SERVICING
The Extreme Shaker is built for long, trouble-free, dependable service. No lubrication
or other technical user maintenance is required. It needs no user maintenance beyond
keeping the surfaces clean. However at least every three (3) months you should:
• Unplug the unit.
• Remove any accumulated dirt from the base and tray.
• Check all accessible items to make sure they are properly tightened.
The unit should be given the care normally required for any electrical appliance. Avoid
wetting or unnecessary exposure to fumes. DO NOT use a cleaning agent or solvent
on the front panel which is abrasive or harmful to plastics, nor one which is flammable.
Always ensure the power is disconnected from the unit prior to any cleaning. If the unit
ever requires service, contact your Ohaus representative.

INTENDED USE

The Ohaus Extreme Shaker are intended for general laboratory use. Safety cannot be
guaranteed if used outside of the intended use.

ENVIRONMENTAL CONDITIONS

Operating Conditions: Indoor use only.
*For use in CO2 environments, incubators, or cold rooms.
Temperature (shaker):
-10 to 60°C (14 to 140°F)
Temperature (controller): -10 to 50°C (14 to 122°F)
Humidity:
up to 100%, condensing
Altitude:
0 to 6,562 ft (2000 M) above sea level
Mains supply voltage:
Fluctuations are not to exceed 10 percent of the
nominal supply voltage.
Non-Operating Storage:
Temperature:
-20 to 65°C (-4 to 149°F)
Humidity:
maximum 80% relative humidity, non-condensing
Installation Category II and Pollution Degree 2 in accordance with IEC 664.
* Avoid cold starts: Unit is not designed to start after being in a cold room environment.
Bring unit into cold room from a room temperature environment, operate and remove
unit from cold room as soon as operation is complete.

EQUIPMENT DISPOSAL

This equipment must not be disposed of with unsorted waste. It is your responsibility to
correctly dispose of the equipment at life-cycle-end by handing it over to
an authorized facility for separate collection and recycling. It is also your
responsibility to decontaminate the equipment in case of biological, chemi-
cal and/or radiological contamination, so as to protect the persons involved
in the disposal and recycling of the equipment from health hazards.
For more information about where you can drop off your waste of equip-
ment, please contact your local dealer from whom you originally purchased this equip-
ment. By doing so, you will help to conserve natural and environmental resources and
you will ensure that your equipment is recycled in a manner that protects human health.
EN
2