4 Adjusting maximum grip force (Fig. 9, Fig. 10)
Pull on System Inner Hand. Use dial 1 (Fig. 9 and Fig. 10, item 1) to set
the power shut-off.
1) Attach 757P5 Test Plug to 757T1 System Tester and to the System
Electric Hand.
2) Simulate opening and closing the System Electric Hand with the 757T1
System Tester.
Note:
The open and close modes of the System Electric Hand must match
the open and close signals on the 757T1 System Tester.
3) Adjust dial 1 so that a maximum grip force of approximately 90 N is
reached. Use the 743F3 Pinch Gauge for checking and adjusting grip force.
5 Setting the upper threshold (Fig. 9, Fig. 10)
Set dial 2 (Fig. 9 and 10, item 2) to the same value as on the dissassembled
Electronic Control Unit.
Use the Ottobock 757T13 MyoSelect to reset the system electric hand to
operation with the 757B15 X-ChangePack. Seal System Inner Hand with
633F11 Silicone Grease and fix with the lock ring.
For further details about the operating modes of the Digital Twin Control,
please refer to the 647H327 Instructions for Use for the System Electric
Hand with Digital Twin Control.
These products may not be disposed of with household waste
in some jurisdictions. Disposal that is not in accordance with
the regulations of your country may have a detrimental impact
on health and the environment. Please observe the information
provided by the responsible authorities in your country regarding
return and collection processes.
6 Liability
Otto Bock Healthcare Products GmbH, hereafter referred to as manufacturer,
assumes liability only if the user complies with the processing, operating
and maintenance instructions as well as the service intervals. The manufac-
turer explicitly states that this product may only be used in combination with
10