8. Fixing the tare weight in position
Fix the tare weight in position using the
knurled nut. Objects on the tray such as
pads etc. weighing up to 3000 g / 6.6 lbs
can be tared using the tare weight.
Model 725

9. Fine adjustment

The hand and counter-hand can be set ex-
actly by turning the fine taring screw. To do
so, the fine taring screw is turned to the
right or left until the hand and counter-hand
line up exactly.
Operation
Once the scale has been prepared as de-
scribed in the section entitled "Set up and
align scale", you can start weighing. Place
the infant in the tray and first slide the large
sliding weight to approximately the right
weight.
Now use the small sliding weight to deter-
mine the exact weight. To do so, slide the
small sliding weight along until hand and
counter-hand line up.
Read the weight value off the scale on the
sliding weight bar.
Cleaning
Clean the housing and the tray as required
using a domestic cleaning agent or com-
mercially-available disinfectant. Follow the
manufacturer's instructions.