Set wavelength
On the pre-test interface, press GOTO λ (2.2) and input the desired wavelength value by using the numeric key-
pad (2.10) (Fig. 8).
Press ENTER (2.7) to save the input value.
Figure 8
Standard curve
0.000
0.000
546.0
Blank
Put the blank sample into the light path and press ZERO (2.3)
Set parameters
Before performing the quantitative analysis several parameters must be set: concentration unit, number of stan-
dard samples to obtain the standard curve and concentration of those standard samples.
On the pre-test interface, press SET (2.1); the instrument will enter the parameter setting display (Fig. 9).
Figure 9
Standard curve
0.000
Abs.
0.000
ml/l
546.0
nm
Select the concentration unit:
- On the parameter setting display, move the cursor with ▼/▲ (2.9) to select Unit and press ENTER (2.7); 8
different units will be displayed (Fig. 10).
- Select the appropriate measurement unit by using the numeric keypad (2.10) or ▼/▲ (2.9) and press
ENTER (2.7).
- If it is not necessary to change the unit press RETURN (2.6).
Figure 10
√ Standard curve
Unit
Number
Concentration
Input the number of standard samples:
- On the parameter setting display, move the cursor with ▼/▲ (2.9) to select Number and press ENTER (2.7)
(Fig. 11).
- By using the numeric keypad (2.10) input the number of standard samples of known concentration that
will be used and press ENTER (2.7). Note: the equipment allows a minimum of 1 standard samples and a maximum
of 9.
Figure 11
√ Standard curve
Unit
Number
Concentration
Input the concentration of standard samples:
- First of all, blank must be performed. Put the blank sample into the light path and, on the pre-test inter-
face, press ZERO (2.3).
- On the parameter setting interface, move the cursor with ▼/▲ (2.9) to select Concentration and press
Page. 24
GOTO λ
Abs.
ml/l
nm
RETURN
√ Standard curve
SET
Unit
Number
Concentration
RETURN
ENTER
RETURN
ENTER
RETURN
Instruction manual 54251050
Standard curve
WL= 546.0 nm
Please Input New:
WL=
√ Standard curve
Display curve
Select curve
1. %
5. ml/l
2. ug/l
6. mg/ml
3.mg/l
7. ug/ml
4. g/l
8. none
Standard curve
N=2
Please Input New:
N=