Verathon GlideScope GO Manual De Funcionamiento Y Mantenimiento página 21

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 86
1.
Ensure the monitor is turned off, and the power adapter or USB flash drive is
disconnected.
2.
Detach the single‑use blade from the monitor, and then dispose of the blade.
3.
Prepare a solution from
4.
Expose the monitor to the solution according to the conditions provided in
5.
Using a towelette or a cotton swab moistened with the solution, clean the power
button, micro‑USB port, the groove around the LCD window, and the groove
where the connector arm attaches to the monitor.
6.
Rinse the monitor according to the conditions provided in
7.
Inspect the monitor for contamination. If any is present, repeat steps 4‑6.
8.
Using a clean cloth, dry the monitor.
9.
Inspect the monitor for damage. If any is present, do not use it, and contact
Verathon
Customer Care.
®
The component should now be clean and free of contamination. Handle the product
carefully to avoid recontamination.
PROCEDURE 2. DISINFECT THE MONITOR
Disinfection of the monitor is optional. Your medical care facility or provider may require
disinfection prior to use. Verathon has validated the solutions and method below for
compatibility and efficacy. For information about additional solutions that may be
available, please contact Verathon Customer Care.
Table 5.
Disinfection Solutions for GlideScope Go
Solution
ASP
CIDEX
OPA
®
®
PDI
Sani‑Cloth
®
®
AF3 Germicidal
Disposable Wipe
1.
Ensure the monitor has been cleaned, according to the prior procedure.
2.
Prepare a solution from
Table 4
according to the manufacturer's instructions.
Level
Up to 1500 cycles as per the following instructions:
Conditioning: N/A
Water temperature: Room temperature
Exposure: Soak the component in CIDEX
High
minutes, ensuring that all air bubbles are removed from the
surface.
Rinse: (3) 1‑minute immersions with agitation in pure water.
Ensure the HDMI connector and micro USB connector are
properly rinsed.
Up to 1500 cycles as per the following instructions:
Conditioning: N/A
Water temperature: N/A
Low
Exposure: Using fresh towelette(s), wet all surfaces and allow
them to remain wet for 3 minutes, paying special attention to
the area around the hinge, all surface contours, and edges.
Rinse: N/A. Allow the component to thoroughly air dry.
Table 5
according to the manufacturer's instructions.
Table
4.
Conditions
OPA for 12
®
Table
4.
15
Tabla de contenido
loading

Tabla de contenido