Positioning
Ensure the surface is flat in the area where the U-Pallet scale will be positioned, especially in the region of the load cell feet.
Small differences in height can be compensated by adjusting the load cell feet.
1. If an Indicator is included with the U-Pallet scale ensure that it is kept with the scale during positioning as there may be
an attached connecting cable.
2. Position the scale at the work site by lifting at the rear handle and rolling it on the two front wheels.
3. Check that the scale surface is level; check that the 4 load cell feet are in contact with the floor; make adjustments by
rotating the nuts on the load cell feet. Use an external level.
4. Ensure that the scale's output cable does not become trapped or crushed during lifting or positioning.
Connecting an Indicator
Certain U-Pallet Scale systems will have an Indicator already attached to the cable end. If the Indicator is not attached or if the
Indicator is purchased separately follow the steps listed below to make the proper connections.
1. Route the cable to the position where the Indicator will be
mounted.
2. Consult the installation instructions of the Indicator to
properly attach the cable connections inside the Indicator.
3. The Signal names of the U-Pallet Scale wires are shown
in Table 1, U-Pallet Scale Wiring Color Code.
CAUTION: Ensure that the cable is protected against possible
damage when routed to the Indicator.
TABLE 1. U-PALLET SCALE WIRING COLOR CODE
Indicator Signal Name
EXC+ (IN+)
SEN+
SIG+ (OUT+)
SIG- (OUT-)
SEN-
EXC- (IN-)
Shield (GND)
VE Series U-Pallet Scale
U-Pallet Scale Wire Color
Green
Yellow
White
Red
Blue
Black
Yellow/Green